Radiologic Technology: Fluoroscopy and Special Procedures

Fluoroscopy equipment, dose reduction technique, and common special procedure workflows tested on radiologic technology certification exams.

34 cards · basic cards · AI-written, checked twice. Edit anything.

Study this set free Look inside first Get FlashKeepers for iPhone
What device converts the x-ray image into visible light in older fluoroscopy units?
The image intensifier
What type of detector has largely replaced the image intensifier in modern fluoroscopy units?
The flat panel detector
What automatic system adjusts kVp and/or mA during fluoroscopy to maintain consistent image brightness?
Automatic brightness control (ABC), also called automatic dose rate control
How does switching to magnification mode affect patient entrance skin dose during fluoroscopy?
It increases patient dose
For a fixed (non-mobile) fluoroscopy unit, what is the minimum source-to-skin distance (SSD) required?
38 cm (15 inches)
For a mobile fluoroscopy unit (C-arm), what is the minimum source-to-skin distance (SSD) required?
30 cm (12 inches)
The fluoroscopy cumulative timer sounds an audible alarm after ____ minutes of beam-on exposure time.
5
What fluoroscopic feature freezes and displays the last acquired image after the exposure stops, reducing the need for additional fluoro time?
Last image hold
How does pulsed fluoroscopy (compared to continuous fluoroscopy) affect patient dose?
It reduces patient dose
What simple technique reduces both patient dose and scatter radiation while improving image contrast during fluoroscopy?
Collimation (restricting the field size)
What effect does increasing distance between the x-ray tube and the patient have on skin dose, according to the inverse square law?
It decreases skin dose
What activation of a special high-level fluoroscopy mode allows exposure rates above the normal limit for better image detail?
Boost mode (high-dose control mode)
During fluoroscopy, where is the patient's entrance skin dose highest?
At the point closest to the x-ray tube
In an under-table (under-couch) fluoroscopy tube configuration, where does most scatter radiation originate relative to the tabletop?
Below the tabletop, attenuated somewhat by the table and patient before reaching the operator
Why does an over-table fluoroscopy tube configuration expose staff to more scatter than an under-table configuration?
The x-ray tube is closer to the operator, and scatter originates near tabletop height without the table attenuating it first