Anesthesia equipment failure troubleshooting and the anesthesia technician's role during an intraoperative emergency tested on the Cer.A.T. certification exam.
About this set
29 basic cards for Cer.A.T. Anesthesia Technology. Set 2 of 10 in Anesthesia Technology. It opens with "What is the anesthesia technician's first action when the oxygen supply failure" and "What safety device automatically cuts off nitrous oxide and other gas flow when". More Healthcare and Nursing decks are listed below.
29 cards · basic cards · AI-written, checked twice. Edit anything.
- What is the anesthesia technician's first action when the oxygen supply failure alarm sounds?
Show answer
Immediately open the emergency backup oxygen cylinder and verify oxygen flow to the patient.
- What safety device automatically cuts off nitrous oxide and other gas flow when oxygen pressure drops below a threshold?
Show answer
The oxygen failure protection device (fail-safe valve).
- What color are medical oxygen cylinders in the United States?
Show answer
Green.
- What color are nitrous oxide cylinders in the United States?
Show answer
Blue.
- What safety system prevents an oxygen regulator from being attached to the wrong gas cylinder?
Show answer
The pin index safety system (PISS).
- What safety system prevents pipeline hoses of different gases from being connected to the wrong wall outlet?
Show answer
The diameter index safety system (DISS).
- Approximately what pressure does a full E-cylinder of oxygen read?
Show answer
About 1900 to 2200 psi.
- What color change in soda lime CO2 absorbent indicates it is exhausted and needs replacement?
Show answer
It turns violet (purple).
- What is the recommended action if CO2 absorbent is suspected of being desiccated?
Show answer
Replace the absorbent before use.
- A rising inspired CO2 reading on capnography during a case suggests failure of what circle system component?
Show answer
A unidirectional (one-way) valve.
- What is the most common cause of a sudden loss of the capnography waveform during a case?
Show answer
Esophageal intubation or a circuit disconnection.
- What test checks the anesthesia machine's breathing circuit for leaks by pressurizing it to about 30 cm H2O with the APL valve closed?
Show answer
The positive pressure leak test.
- What test checks the anesthesia machine for leaks between the flowmeters and the common gas outlet?
Show answer
The negative pressure leak test.
- What part of the anesthesia machine does the high pressure leak test check?
Show answer
The system from the gas cylinder to the first-stage pressure regulator.
- What happens if the APL (adjustable pressure limiting) valve becomes stuck fully closed during manual ventilation?
Show answer
Circuit pressure rises rapidly, risking barotrauma to the patient.